Ethos® Automated Ultrasound
Probe Cleaner Disinfector

How Ethos Works

1

Prepare a soiled ultrasound probe for Ethos by removing the procedure cover with a wipe. The technician does not need to perform any other manual steps.

Step 1 - Insert the Probe

Top: A soiled ultrasound probe after a procedure, with the procedure cover still in place.

Bottom: The procedure cover has been removed and the soiled transducer is now ready to place in Ethos.

2

Open the probe door on Ethos and place the soiled ultrasound probe into the cleaning and disinfection chamber as indicated on the LCD screen.

Step 2 - Automated Cleaning

Open the probe door and insert the soiled transducer, following the on-screen instructions.

3

Once the probe is placed in the reprocessing chamber of Ethos, a tempered water spray activates to remove any gross matter from the probe surface. During this time Ethos is mixing the cleaner disinfectant, AquaCide®, for application onto the surface of the soiled ultrasound probe. AquaCide is mixed and automatically tested for MRC before being sprayed onto the surface of the probe with a coverage area ranging from the distal tip to the strain relief. AquaCide performs both cleaning and high-level disinfection in 3 minutes.

4

At the conclusion of the cleaning and high-level disinfection, Ethos then sprays the entire probe with 5-nanometer filtered water to remove any residue from the cleaning and HLD application.

Step 4 - Rinse and Dry

The probe is bathed in filtered rinse-water, ensuring no residue remains.

5

Remove the cleaned and high-level disinfected probe from Ethos. Dry the probe using a contaminant-free drying cloth, such as QwikDry®, and then store it in a HEPA-filtered storage cabinet or transport it in an approved device, like Probe Valet™, back to the procedure area for subsequent patient use.

Step 5 - Probe Ready for Use

Remove the probe from Ethos and dry it using a contaminant-free cloth.